BBS-EX

Extended Area Blackbody Calibration Source

The PSC EX series of black-body sources offers a range of temperatures in a 12″ x 12″ black-body calibration source. Ambient temperature ranges from 230 °C to 500°C.

The BBS-EX Series (BBS230-EX, BBS350-EX, and BBS500-EX) are high-performance extended area blackbody calibration sources designed for accurate calibration of infrared measurement systems over large target areas. These systems provide a uniform radiating surface, making them ideal for calibrating thermal imaging cameras, infrared thermometers, and large-field sensors in industrial, laboratory, and research environments.

With a large 12” × 12” emitter surface, the BBS-EX series ensures consistent and uniform radiation across wide measurement fields. Each model offers a different temperature range (up to 230°C, 350°C, or 500°C), allowing users to select the most suitable solution for their specific application requirements.

Engineered for precision, the series features high emissivity (0.96 ± 0.02), excellent stability (±0.1°C), and NIST-traceable accuracy, ensuring reliable and repeatable calibration results. The advanced multi-band PID control system maintains precise temperature regulation, while flexible communication interfaces (RS-232 and GPIB) enable easy system integration.

Built with a robust design and fast heat-up capability, the BBS-EX series delivers dependable performance for demanding calibration tasks, making it an ideal solution for large-area infrared calibration and thermal imaging system validation.

BBS230-EX BBS350-EX BBS500-EX
Product Type Extended Area Blackbody Source Extended Area Blackbody Source Extended Area Blackbody Source
Temperature Range Ambient to 230°C Ambient to 350°C Ambient to 500°C
Wavelength Range 1 – 99 µm 1 – 99 µm 1 – 99 µm
Emissivity 0.96 ± 0.02 0.96 ± 0.02 0.96 ± 0.02
Emitter Size 12” × 12” (304 × 304 mm) Same Same
Source Type Extended Area Extended Area Extended Area
Temperature Resolution 0.1°C 0.1°C 0.1°C
Calibration Accuracy ±0.2°C + 1 digit (NIST) ±0.2°C + 1 digit (NIST) ±0.2°C + 1 digit (NIST)
Stability ±0.1°C ±0.1°C ±0.1°C
Response Time < 50 minutes < 50 minutes < 50 minutes
Temperature Sensors Platinum RTD & Type T Platinum RTD & Type T Platinum RTD & Type S
Control Type Active Multi-Band PID Active Multi-Band PID Active Multi-Band PID
Power Requirements 1500 W max 2000 W max 4500 W max
Voltage 90–125 / 208–240 VAC 90–125 / 208–240 VAC 208–240 VAC
Interface RS-232 / IEEE-488 (GPIB) Same Same
Dimensions (Source) 17” × 17” × 17” Same Same
Controller Dimensions 5.1” × 13.4” × 12” Same Same
Weight (Emitter) 50 lbs 50 lbs 60 lbs
Weight (Controller) 9 lbs 9 lbs 9 lbs